Please use this identifier to cite or link to this item:
http://ricaxcan.uaz.edu.mx/jspui/handle/20.500.11845/733| Title: | Passive neutron area monitor with pairs of TLDs as neutron detector |
| Authors: | Vega Carrillo, Héctor René Guzmán García, Karen Arlete Gallego Díaz, Eduardo Lorente, Alfredo |
| Issue Date: | Oct-2014 |
| Publisher: | Elsevier |
| Abstract: | A passive neutron area monitor has been designed using Monte Carlo methods; the monitor is a polyethylene cylinder with pairs of thermoluminescent dosimeters (TLD600 and TLD700) as thermal neutron detector. The monitor was calibrated with a bare and a thermalzed 241AmBe neutron sources and its performance was evaluated measuring the ambient dose equivalent due to photoneutrons produced by a 15 MV linear accelerator for radiotherapy and the neutrons in the output of a TRIGA Mark III radial beam port. |
| URI: | http://localhost/xmlui/handle/20.500.11845/733 https://doi.org/10.48779/05zk-zg42 |
| ISSN: | 1350-4487 |
| Other Identifiers: | info:eu-repo/semantics/publishedVersion |
| Appears in Collections: | *Documentos Académicos*-- UA Ciencias Nucleares |
Files in This Item:
| File | Description | Size | Format | |
|---|---|---|---|---|
| Passive neutron area.pdf | 3,18 MB | Adobe PDF | ![]() View/Open |